| Model Number |
1203350-38 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Difficult to Advance (2920); Material Deformation (2976); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion in an unspecified artery.The 3.5x38mm esprit btk would not advance onto the 0.014" command es guidewire.The device did not enter sheath and would not go further.Scaffold and balloon were noted to be bunched up and deformed.Then the esprit could not be removed from the guidewire either.The guidewire was cut just before the balloon tip to proceed.The same guidewire with another esprit was used with no issue to complete the procedure.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.C3: 1 per year was entered; however, the frequency is 1 dose per implant.The additional device referenced in b5 & d10 is filed under separate medwatch report number.
|
| |
|