MAUDE Adverse Event Report: ABBOTT VASCULAR INC. Esprit¿; Scaffold,infrapopliteal,absorbable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR INC. Esprit¿; Scaffold,infrapopliteal,absorbable Back to Search Results
Model Number 1203350-38
Medical Device Problem Codes Difficult to Remove (1528); Difficult to Advance (2920); Material Deformation (2976); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the procedure was to treat a lesion in an unspecified artery.The 3.5x38mm esprit btk would not advance onto the 0.014" command es guidewire.The device did not enter sheath and would not go further.Scaffold and balloon were noted to be bunched up and deformed.Then the esprit could not be removed from the guidewire either.The guidewire was cut just before the balloon tip to proceed.The same guidewire with another esprit was used with no issue to complete the procedure.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.C3: 1 per year was entered; however, the frequency is 1 dose per implant.The additional device referenced in b5 & d10 is filed under separate medwatch report number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Esprit¿
Common Device Name
Scaffold,infrapopliteal,absorbable
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 -280
Manufacturer (Section G)
ABBOTT VASCULAR, Reg # 2024168
26531 ynez road
temecula CA 92591 -462
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591--462
+1(951)914-5259
MDR Report Key50758254
Report Number2024168-2026-03531
Device Sequence Number18440469
Product Code NXW
UDI-Device Identifier08717648357091
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P230036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1203350-38
Device Catalogue Number1203350-38
Device Lot Number5101461
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2026
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/14/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
0.014" Command ES
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-