| Model Number |
863283 |
| Medical Device Problem Code |
Signal Artifact/Noise (1036)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint on the suresigns vs4 nbp, spo2 indicating that the devices o2 monitor has a ghost read.It is unknown if the device was in use at time of event, and there was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A philips remote service engineer (rse) spoke with the biomed regarding the device's inaccurate spo2 parameter.The issue could not be replicated, and the cause is unknown.The rse advised the biomed that bench repair was not available because the device is eol/eos.The system does not meet the specification for the performed service and is not returned to use.The device does not meet the specification for the performed service and is not returned to use.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
| |
|