| Catalog Number |
043.161S |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Codes |
Pain (1994); Fibrosis (3167)
|
| Date of Event |
07/28/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The batch number could be verified.Our manufacturing q-system assures,\
that production and process controls are in place to ensure that batches\
confirm to the applicable specifications before they are distributed.?\
the loss of an endosseous dental implant after successful\
osseointegration and restoration is a known inherent risk of the\
procedure.\
\
implants may have to be removed in case one or more of the implant\
success criteria are not met.\
implant success criteria according to buser et al.(1991) are:\
1.Absence of persistent subjective complaints such as pain, foreign\
body sensation and /or dysesthesia\
2.Absence of a recurrent peri-implant infection with suppuration\
3.Absence of implant mobility\
4.Absence of a continuous radiolucency around the implant.\
\
the manufacturer's trend analysis confirms that the reported failure\
rate associated with its dental implants is below the expected failure\
rate for this treatment as published in the scientific literature.
|
| |
|
Event or Problem Description
|
|
The clinician reports the implant was inserted 2021(b)(6) in fdi 36.On 2026(b)(6), loss of osseointegration was verified.The device was forwarded to the manufacturer.At the event the patient experienced: pain and mobility.No further patient complications were reported.
|
| |
|