The account reported finding defective cautery and that the cautery would not work during the procedure; there was no report of any patient injury or adverse event.A sample was received for investigation, and medline received one used cautery pencil for sample review; the pencil tip had evidence of burning, the buttons clicked and did not stick, and there were no visual defects with the chord or plugs.The investigation and sample evaluation confirmed the reported issue.Medline determined that the root cause is a vendor product issue related to the component manufacturing process, and the vendor confirmed the reported issue through replication and determined that the device identified intermittent activation; the complaint was provided to the vendor for further evaluation and supplier quality will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|