| Model Number |
GF-UCT180 |
| Medical Device Problem Codes |
Device Reprocessing Problem (1091); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event is unknown.Additional information will be provided once available.\
\
the final investigation has been completed.The failures identified during the inspection have been confirmed.\
\
a root cause for the foreign object events could not be identified.Based on the results of the investigation, the most likely causes were not established.\
\
a root cause for the chipped adhesive event could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: expected or random component failure without any design or manufacturing issue.\
\
olympus will continue to monitor the field performance of this device.\
\
should any additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer returned the ultrasound gastrovideoscope to the service center for a separate issue.During the device analysis, the service repair technician indicated that the following malfunctions were found: an objective lens cleaning failure, distal end residue, a chip in the adhesive area between the distal end and the light guide cover, and the forceps elevator pipe was clogged with foreign substances.There was no patient involvement.
|
| |
|