Abbott Medical TriClip G5 Delivery System; tricuspid valve repair device, percutaneously delivered Back to Search Results
Model Number TCDS0801-XTW
Medical Device Problem Code Incomplete Coaptation (2507)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2026, a patient presented with grade 4 functional tricuspid regurgitation (tr) for a triclip procedure.During the procedure, first triclip was implanted on medial side of anterior and septal leaflet (a/s).Post deployment, it was determined that there was a single leaflet device attachment (slda) and the clip was no longer attached to the anterior leaflet.Second triclip xt and third triclip xtw was implanted to stabilize the slda.The final tr was grade 1-2.There were no adverse patient effects or clinically significant delays reported.
 
Event or Problem Description
It was reported that on (b)(6) 2026, a patient presented with grade 4 functional tricuspid regurgitation (tr) for a triclip procedure.During the procedure, first triclip was implanted on medial side of anterior and septal leaflet (a/s).Post deployment, it was determined that there was a single leaflet device attachment (slda) and the clip was no longer attached to the anterior leaflet.Second triclip xt and third triclip xtw was implanted to stabilize the slda.The final tr was grade 1-2.There were no adverse patient effects or clinically significant delays reported.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.Based on available information, the cause of the reported slda was unable to be determined.The reported unexpected medical intervention was the result of case-specific circumstances.
 
Additional Manufacturer Narrative
The device remains implanted in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TriClip G5 Delivery System
Common Device Name
tricuspid valve repair device, percutaneously delivered
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, Reg # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50762752
Report Number2135147-2026-06217
Device Sequence Number9668799
Product Code NPS
UDI-Device Identifier05415067050731
UDI-Public(01)05415067050731(17)271008(10)60408R1118
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P230007
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTCDS0801-XTW
Device Catalogue NumberTCDS0801-XTW
Device Lot Number60408R1118
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2026
Supplement Date Received by Manufacturer09/15/2026
Initial Report FDA Received Date09/14/2026
Supplement Report FDA Received Date10/01/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/10/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Triclip steerable guide catheter
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexFemale
-
-