| Model Number |
TCDS0801-XTW |
| Medical Device Problem Code |
Incomplete Coaptation (2507)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a patient presented with grade 4 functional tricuspid regurgitation (tr) for a triclip procedure.During the procedure, first triclip was implanted on medial side of anterior and septal leaflet (a/s).Post deployment, it was determined that there was a single leaflet device attachment (slda) and the clip was no longer attached to the anterior leaflet.Second triclip xt and third triclip xtw was implanted to stabilize the slda.The final tr was grade 1-2.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a patient presented with grade 4 functional tricuspid regurgitation (tr) for a triclip procedure.During the procedure, first triclip was implanted on medial side of anterior and septal leaflet (a/s).Post deployment, it was determined that there was a single leaflet device attachment (slda) and the clip was no longer attached to the anterior leaflet.Second triclip xt and third triclip xtw was implanted to stabilize the slda.The final tr was grade 1-2.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.Based on available information, the cause of the reported slda was unable to be determined.The reported unexpected medical intervention was the result of case-specific circumstances.
|
| |
|
Additional Manufacturer Narrative
|
|
The device remains implanted in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|