The account reported finding betadine leaked in pack.A photo and a sample were received for investigation; four unopen packs were received, pvp paint leaked in each of the four packs, there was no damage to the pvp bottles, production records had no non-conformances, and there were no issues with the pack build that might have an impact on the issue.The reported problem was confirmed based on the sample evaluation.Based on the component's location in the pack it is unlikely that placement contributed to the reported issue, and this complaint has been provided to the vendor for further evaluation while the supplier quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|