The account reported finding cautery burning off during use.A sample was received for the investigation, and medline received one used suction coagulator; the component received had a blue plastic body which does not match the suction coagulator in this kit¿s bom, and the insulation on the tip end does appear to have been burned slightly.Medline was unable to confirm the reported issue with the sample received as the received sample is not the suction coagulator that is in this kit's pack build; the cause was documented as unconfirmed defect, sample required to determine cause.The complaint has been provided to the vendor for further evaluation, supplier quality will monitor this issue for trending purposes, and no medical intervention, serious injury, or follow-up care has been reported.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|