MAUDE Adverse Event Report: Biosense Webster Inc OPTRELL; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Biosense Webster Inc OPTRELL; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Catalog Number D140901
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Cardiac Tamponade (2226)
Date of Event 08/18/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An analysis of the product could not be performed since a physical sample was not received for evaluation.\ \ a manufacturing record evaluation was performed for the finished device number lot 31274523m and no internal actions related to the complaint were found during the review.\ \ as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.\ \ based on the information available, there is no indication that a design or manufacturing issue has caused the reported complaint condition.\ \ if the product is received at a later date, the investigation will be updated as applicable.\ \ this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.\ \ manufacturer's reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient underwent an ablation procedure with an optrell catheter and the patient experienced cardiac tamponade.\ \ after premature ventricular contraction procedure, approximately one hour after the optrell use, during the brockenbrough method (a transseptal puncture technique) was about to be performed, pericardial effusion and a decrease in blood pressure were confirmed.There was no malfunction or breakage of the biosense webster inc.(bwi) product itself.Since no bwi product malfunction was observed, after the unknown treatment for tamponade, the procedure was continued and completed without issue.The patient¿s outcome has improved.\ \ physician assessment of the health hazard is non-serious (moderate/minor).The physician¿s opinion on the cause of the event is that it is procedure related.The physician commented that the issue might have occurred either due to the aiv catheter wire or while advancing the optrell catheter into the right ventricular outflow tract.There were no abnormalities observed prior to or during use of the product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTRELL
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
Biosense Webster Inc
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
Biosense Webster Inc (Juarez)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
katie seppa
31 technology dr
irvine, CA 92618
+1(513)337-8052
MDR Report Key50763117
Report Number3013300026-2026-03013
Device Sequence Number21172612
Product Code DRF
UDI-Device Identifier10846835023015
UDI-Public(01)10846835023015(17)270314(10)31274523M
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K211438
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberD140901
Device Lot Number31274523M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2026
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/15/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexFemale
Patient RaceAsian
-
-