• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Ireland Diagnostics Division Alinity i TSH Reagent Kit; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Ireland Diagnostics Division Alinity i TSH Reagent Kit; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE Back to Search Results
Catalog Number 07P48-30
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Alinity i TSH Reagent Kit
Common Device Name
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
Abbott Ireland Diagnostics Division
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
Abbott Ireland Diagnostics Division
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
+49()6122582960
MDR Report Key50763314
Report Number3005094123-2026-00454
Device Sequence Number15806405
Product Code JLW
UDI-Device Identifier00380740131159
UDI-Public(01)00380740131159(17)270911(10)86104UD00
Combination Product (Y/N)N
Initial Reporter CountrySA
PMA/510(K) Number
K983442
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number07P48-30
Device Lot Number86104UD00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2026
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/14/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age36 YR
Patient SexFemale
-
-