The complaint description was splitting/torn in the middle, and the customer stated they have 2 boxes of gloves that are splitting/torn in the middle; the issue occur during use was answered no and injury was answered no.The account provided 2 used boxes of gloves to support the investigation, and thirteen gloves were donned and evaluated, including measurements of glove thickness, palm width, and overall length from the tip of the middle finger to the cuff.During testing, no glove tore and all measured dimension remained within specifications; based on the testing performed and the measurement results, medline was not able to confirm the reported issue.Medline was unable to replicate or confirm the issue reported with the provided sample, and as a result, could not identify the root cause; there were no reports of death or serious injury associated with this event.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|