• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical 8CH Infinity DBS Flex Extn Kit, 60cm, B; DBS Lead Extension

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical 8CH Infinity DBS Flex Extn Kit, 60cm, B; DBS Lead Extension Back to Search Results
Model Number 6372
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient's system was unable to enter mri mode.Troubleshooting revealed high impedances on the right lead.As a result, surgical intervention was undertaken on (b)(6)2026 wherein the lead extension was replaced to address the issue.Therapy was restored post-operatively.
 
Additional Manufacturer Narrative
Date of event is estimated.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8CH Infinity DBS Flex Extn Kit, 60cm, B
Common Device Name
DBS Lead Extension
Manufacturer (Section D)
Abbott Medical
6901 preston rd
plano TX 75024
Manufacturer (Section G)
Abbott Medical
6901 preston rd
plano TX 75024
Manufacturer Contact
sara speyerer
6901 preston road
plano, TX 75024
+1(972)309-8000
MDR Report Key50763996
Report Number1627487-2026-10050
Device Sequence Number13178728
Product Code MHY
UDI-Device Identifier05415067030351
UDI-Public(01)05415067030351(10)8651321(17)240906
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P140009
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/06/2024
Device Model Number6372
Device Catalogue Number6372
Device Lot Number8651321
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2026
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DBS Burr Hole Cap; DBS IPG; DBS Lead
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient Weight245 LB
-
-