The customer reported that the nurse pricked his finger with needle when removing from packaging due to clear cap not on needle; the customer also stated the needle stick occurred on the left ring finger and was deep enough to make bleed, with no medical intervention required, no follow-up care needed, no post-exposure tasks, and no impact to the patient/provider.No samples were sent in for investigation, but a lot number was recorded; reserve stock from the same lot was ordered, and 25 samples were randomly selected for visual and functional testing.Visual inspection revealed no manufacturing defects preventing the samples to work properly, functional testing was unable to create the same issue, and the caps stayed firmly on the needle and showed no signs of falling off during the opening process or when being attached to the syringe; the complaint was not confirmed.The cause was recorded as unconfirmed defect - not confirmed with received sample, material trending via sap showed no similar complaints during the last 12 months, and division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|