| Model Number |
EGIAUXL |
| Medical Device Problem Codes |
Component or Accessory Incompatibility (2897); Mechanical Jam (2983); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during a procedure, prior to patient use, the device jammed and the reload was locked and unable to be loaded.A new product was used to resolve the issue.There was no patient injury.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a procedure, prior to patient use, the reload got stuck and unable to be loaded.A new reload was used to resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant product: egia45avm egia45avm egia 45 artic vasc med sulu (lot #p5j1162).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b3, d1, d4, d8, g3, g4, h4, h6 further review of this event found the device is not considered a potential contributory cause.This event has been reassessed as no n-reportable and no further reports will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|