Catalog Number 533-06-108
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Periprosthetic fracture (4902)
Date of Event 08/24/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to periprosthetic fracture.
 
Additional Manufacturer Narrative
The agent reported: "patient was converted from reverse to revision reverse with allograft for a periprosthetic fx.The original baseplate, 2 peripheral screws, and glenopshere were retained.The stem, poly, and x2 peripheral screws were removed for this case.We were unable to locate the original d ticket.Female 81".The previous surgery and the surgery detailed in this event occurred 7 years, 10 months, 26 days apart.Item number: 533-06-108 item description: altivate reverse, humeral stem, small shell, sz 6x108mm lot#: 923w1051 part revision: a product type: shoulder manufacture date:04-sep-2018 expiration date: 27-aug-2024.This evaluation is limited in scope as the item(s) associated with this investigation was not returned to djo surgical - austin for examination.The surgery was completed as intended.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate that the reported device(s) was defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.Root cause: the root cause of this complaint was a periprosthetic fracture requiring conversion from a reverse shoulder arthroplasty to a revision reverse shoulder arthroplasty with allograft reconstruction.There are multiple factors that may contribute to an event that are outside of the control of enovis surgical.Containment: inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.Device history records review a review of the device history record(s) shows that the reported component(s) used in the previous surgery met design and manufacturing requirements at the time of release for use.There were no nonconforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Complaint history: customer complaint history of the reported device(s) showed no present trends or ongoing issues that need review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
ALTIVATE REVERSE, HUMERAL STEM, SMALL SHELL, SZ 6X108MM
Manufacturer (Section D)
Encore Medical L.P
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
Encore Medical L.P
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd.
austin, TX 78758
MDR Report Key50766719
Report Number1644408-2026-01841
Device Sequence Number21249746
Product Code KWS
UDI-Device Identifier00190446259091
UDI-Public00190446259091
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K172351
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/27/2026
Device Catalogue Number533-06-108
Device Lot Number923W1051
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2026
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/04/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
506-03-130 Lot. 835C1387; 506-03-130 Lot. 835C1434; 509-02-032 Lot. 951W1067
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
Patient SexFemale
-
-