Bard Peripheral Vascular, Inc. True Dilatation Catheter; Balloon Valvuloplasty Catheter
|
Back to Search Results |
|
| Catalog Number |
0234512 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/06/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6), 2026, a patient underwent a valvuloplasty procedure using a true balloon catheter.During the procedure, the balloon catheter could not be retracted into the sheath as intended, resulting in difficulty removing the sheath.To facilitate device retrieval, the balloon catheter was snared through a secondary sheath via contralateral femoral access and externalized through the left femoral access site.The balloon catheter was subsequently cut, and the remaining catheter shaft was extracted through the rfa sheath.The procedure was completed using another device.The patient's current status is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h.4.Device manufacture date: unknown
as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|
|
|