| Model Number |
CF-HQ190I |
| Medical Device Problem Code |
Image Display Error/Artifact (1304)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the colonovideoscope had image disturbances during reprocessing.During the initial evaluation of the device it was observed that the image disappeared.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The final investigation has been completed.The failure reported by the customer and identified during the inspection have been confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunctions: due to damage on charged coupled device unit, the image disappeared during angulation in right and left directions.The cause was determined to be related to a component failure.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post-market surveillance activities.
|
| |
|