| Model Number |
470230-12 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, a large hem-o-lok clip applier was not releasing the clips appropriately.There was no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The large hem-o clip applier has been returned and evaluated by the failure analysis team.Failure analysis confirmed and replicated the reported complaint that the clip applier was not clipping.Investigation identified one bent and misaligned grip tip without cracks, which caused the instrument to fail the clip test due to improper grasping, failure to hold clips, and release difficulties.The complaint was confirmed by failure analysis.
|
| |
|
Search Alerts/Recalls
|