NOVOCURE GmbH Optune Gio Back to Search Results
Model Number TFH9100
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Wound Dehiscence (1154)
Date of Event 06/23/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A 39-year-old male patient with newly diagnosed glioblastoma (gbm) started optune gio therapy on (b)(6) 2026.On (b)(6) 2026, novocure received a medical record dated (b)(6) 2026, indicating that the patient was experiencing mild irritation at the surgical incision site, which remained slightly open.During a follow up appointment on (b)(6) 2026, it was noted that the patient tolerated optune gio well without skin side effects.An attempt was made to contact the prescribing physician; however, no reply was received.
 
Additional Manufacturer Narrative
Based on the available information, the contribution of the array placement to the wound dehiscence that might lead to a serious event, requiring medical and/or surgical intervention cannot be ruled out.Wound dehiscence is an expected event with optune gio device use (ef-11 0% and <1% ef-14 optune arm).Contributing factors for wound dehiscence in this patient include: prior radiation, underlying cancer disease, prior surgery affecting skin integrity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Optune Gio
Common Device Name
Optune Gio
Manufacturer (Section D)
NOVOCURE GmbH
21 neuhofstrasse
baar 6340
SZ  6340
Manufacturer Contact
sharon perez
180 hanover street
portsmouth, NH 
+1(207)752-7602
MDR Report Key50767402
Report Number3010457505-2026-00857
Device Sequence Number18529436
Product Code NZK
UDI-Device Identifier07290107986427
UDI-Public(01)07290107986427(11)180315
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P100034/S033
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberTFH9100
Device Catalogue NumberTFH9100US
Device Lot NumberN/A
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 08/18/2026
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/15/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age39 YR
Patient SexMale
-
-