| Model Number |
500004502000 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medflex 1000, a user was unable to override a medication for a patient profile.Investigation revealed that the patient did not have an active morphine order in the system.The customer was attempting to add and override the medication; however, the medication was classified as a high alert medication, and the user only had general alert access privileges.As a result, the override could not be completed.The customer reported that a malfunction took place when the user tried to dispense medication to the patient.There were no delay or adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 07-nov-2013 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the user was unable to override a medication to a patient profile.A technical support specialist (tss) review of myql showed that the patient did not have an active morphine order.The customer was attempting to override the medication; however, the medication was classified as high alert and the user only had general alert privileges, preventing the override.The option to contact the pharmacy was offered for further assistance, but the customer ended the call before additional action could be taken.The system functioned as intended after the technical support specialist investigated the device.
|
| |
|
Search Alerts/Recalls
|