MAUDE Adverse Event Report: Medtronic, Inc. C320LBB Delivery System; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic, Inc. C320LBB Delivery System; CATHETER, PERCUTANEOUS Back to Search Results
Model Number C320LBBS45
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the implant procedure, the catheter was performing as expected during the delivery and deployment of the lead.When the physician slit the catheter, it snapped in half but the remnants were extracted and a hemostat was used to continue to slit the sheath.The lead was stable and was successfully implanted.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the delivery catheter was returned and analyzed.There was an issue with the catheter shaft.The catheter shaft was torn.The shaft of the delivery catheter was spiral slit.The shaft on the hub of the delivery catheter was spiral slit.The septum detached from the hub of the delivery catheter.The septum of the catheter was damaged.Visual analysis of the delivery catheter indicated damage during use.The analyst noted the delivery catheter was returned in two pieces.The slitter was not returned.The septum of the delivery catheter was damaged with the adhesive partially releasing from the hub of the delivery catheter.Slitting started near the centerline on the hub of the delivery catheter.The shaft of the delivery catheter was torn in multiple locations along the shaft of the delivery catheter.The shaft of the delivery catheter was torn/broke at 23.6 cm.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C320LBB Delivery System
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
Medtronic, Inc.
8200 coral sea street ne
mounds view,mn
Manufacturer (Section G)
Medtronic, Inc.
8200 coral sea street ne
mounds view,mn
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50767790
Report Number2182208-2026-51838
Device Sequence Number16056416
Product Code DQY
UDI-Device Identifier00199150070138
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K253409
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberC320LBBS45
Device Catalogue NumberC320LBBS45
Device Lot Number013011726A
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/08/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2026
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/22/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexFemale
-
-