A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 22-sep-2025 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the latch required adjustment.A technical support specialist (tss) remoted in and found the drawer higlighted in yellow and not populating in tester, even after a reboot.Field service engineer (fse) performed inspection of the medbank tower and found no active errors.The rear panel was opened, board status lights were verified as operational, and the drawers were inspected.Rack 1 was manually opened, and the red latch lever on drawer 4 was adjusted.The medbank was powered on, and hardware test application (hta) testing was performed on all matrix drawers with passing results saved in the folder.The system was rebooted, and the user successfully inventoried medications in drawer 4.Repair was verified through successful hta testing and medication inventory completion.The system functioned as intended after the field service engineer repaired the device.
|