H3: product analysis :evaluation determined that attachment is not holding onto the burr after being correctly locked.The likely cause was identified as due to distal bearings are detached.It was also noted that input and output drive shafts are worn, with some component stuck on the output shaft due to deformation, tube tip is deformed and the laser markings were faded.G.3.Aware date used as date of analysis performed date as the event is reported based on evaluation findings.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|