A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 01-oct-2013 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the drawer had failed.The field service engineer (fse) reseated the row board cable connections at the drawer controller board and row board trays.The fse inspected the cables for damage, including burn marks, exposed wiring, and cuts, and removed pockets to clear any debris that could have caused electrical shorts.Hardware test application testing was performed, and the drawer function was verified successfully.The system functioned as intended after field service engineer repaired the device.
|