| Model Number |
MR8-AVA12-R |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that bearing detachment was observed prior to use.There is no patient impact reported.
|
| |
|
Event or Problem Description
|
|
It was reported that bearing detachment was observed prior to use.There is no patient impact reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis: no conclusion can be drawn.No evaluation was performed, as the device is returned and analysis is not yet completed.Mdr timeline is coming due but analysis not yet finished.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis :evaluation determined that the inner race of the bearing fell out.The likely cause was identified as bearings internal tubes are corroded.It was also noted that input and output drive shaft are corroded and the laser markings are faded.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|