| Model Number |
0998-00-3013-45 |
| Medical Device Problem Code |
Inability to Auto-Fill (1044)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the customer that during use, the cs100 intra-aortic balloon pump (iabp) malfunctioned.Unit displayed autofill failure alarm.User replaced the unit with another one.Patient was involved.There was no patient harm reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that the fault was caused by insufficient negative pressure leading to an automatic inflation failure, requiring the replacement of the 5000-hour maintenance kit.Parts were replaced.Product passed all functional and safety tests per manufacturer specifications.Unit was delivered to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
E.(b)(6).Upon completion of the investigation into this event, a follow up report will be submitted.
|
| |
|