Xenios AG XLUNG Kit 230; Extracorporeal System For Long-Term Respiratory / Cardiopulmonary Failure Back to Search Results
Catalog Number 32000014
Medical Device Problem Code Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2026
Type of Reportable Event Malfunction
Event or Problem Description
A user facility reported that the circuit was extremely noisy during support.Console information included 38350043, s/n (b)(6) and the pump drive information included 28350048, s/n (b)(6).The speed of the console was set at 5,800 rpm.The kit information included lot gsxl0304.Due to the noise from the pump, and as the patient was able to be weaned from support, the circuit was removed and discarded onsite.There was no indication of patient serious injury.
 
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XLUNG Kit 230
Common Device Name
Extracorporeal System For Long-Term Respiratory / Cardiopulmonary Failure
Manufacturer (Section D)
Xenios AG
director of quality
im zukunftspark 1
heilbronn baden-wurttemberg 74076
GM  74076
Manufacturer (Section G)
Xenios AG
director of quality
im zukunftspark 1
heilbronn baden-wurttemberg 74076
GM   74076
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
+1(617)417-5172
MDR Report Key50770819
Report Number3012172416-2026-00084
Device Sequence Number9472384
Product Code QJZ
UDI-Device Identifier04057224001494
UDI-Public04057224001494
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K191407
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Pharmacist
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number32000014
Device Lot NumberGSXL0304
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2026
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Novalung Console (Linux Board); Novalung Console (Linux Board).
Patient Age40 YR
Patient SexMale
Patient Weight176 LB
-
-