The customer reported one is collapsed, and they have tried to problem solve however they cannot get it to inflate and therefore cannot use.Medline received 1ea bulb assembly and valve in used condition and tested the sample according to the instructions for use; when attempting to inflate the cuff, the inflation bulb would not inflate the cuff as intended, the bulb appeared to be delated, and air could not be put back in the bulb.The complaint was confirmed.Issue shown via investigation was caused by a defective inflation bulb, due to the air mechanism being blocked; the sample did not work as intended, and the division will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|