It was reported that the implant at tooth site #36 was removed due to peri-implantitis.During the impression appointment for crown fabrication, the practitioner noted granulation tissue on the buccal aspect of the implant and probing extending to the apex.Curettage was performed, and bone graft material was placed; however, the condition did not improve, and the implant was subsequently removed.
|
|
Zimvie complaint number (b)(4).A4: patient weight unknown / not provided.E1: reporter email address unknown / not provided.G4: premarket identification k011028|k013227.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|