Medline Industries, LP MEDLINE; LIFT,PAT,HYDRAULIC,6 POINT CRD,NO SLING Back to Search Results
Catalog Number MDS88200HL
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the wheel with the brake on left side is broken and the ball bearings have came out.
 
Additional Manufacturer Narrative
The customer stated the wheel with the brake on left side is broken and the ball bearings have come out, and per customer, the ball bearings came out of the left rear caster.A photo was provided; a physical sample was not provided, so only a visual review of the defect could be conducted, and a functional or dimensional inspection could not be executed via a photo.The investigation results were not confirmed; after reviewing the photo it was discovered that the issue cannot be identified based on the photo provided.No root cause was identified; no product changes were identified that could have contributed to the alleged product failure, quality control inspection records were reviewed, and corrective/preventive action was not taken because the division will initiate further investigation and/or supplier corrective action requests based on trending results.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
LIFT,PAT,HYDRAULIC,6 POINT CRD,NO SLING
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
diana arizmendi
three lakes drive
northfield, IL 60093--275
MDR Report Key50771289
Report Number1417592-2026-40295
Device Sequence Number13133734
Product Code FSA
UDI-Device Identifier40884389144825
UDI-Public40884389144825
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS88200HL
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2023
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-