| Catalog Number |
11097542 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that they obtained erroneously depressed tacrolimus (tacr) results on two patient samples on an atellica ch 930 analyzer.The initial results were reported to the physician(s) and were questioned.The samples were repeated on an alternate atellica ch 930 analyzer, and the results recovered higher than the initial results and were considered correct.Corrected reports were issued with the repeat results.There are no known reports of patient intervention or adverse health consequences due to the erroneously depressed tacr results.
|
| |
|
Additional Manufacturer Narrative
|
|
A united states (us) customer contacted the siemens healthcare diagnostics remote services center (rsc) regarding erroneously depressed tacrolimus (tacr) results obtained on two patient samples on an atellica ch 930 analyzer.Calibration and quality control (qc) recovered acceptably on the event date.Siemens is evaluating the event.
|
| |
|