| Catalog Number |
04T91-20 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported a falsely elevated alinity c creatinine2 result generated on the alinity c processing module.The questioned creatinine result did not match previous results, and the patient was admitted to the hospital.The following data was provided:
(b)(6)2026, sid (b)(6): initial creatinine2 result = 2.37 mg/dl (normal range: 0.72 to 1.25 mg/dl)
(b)(6)2026, repeat creatinine2 result = 0.70 mg/dl (measured from original sample stored in the refrigerator)
there was no impact to patient management reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Section a patient information: refer to section b5 for complete patient information.All available patient information was included.Additional patient details are not available.
|
| |
|