The customer reported issues with the hydrodissectors in the packs: they appear to be crimped and causing a double barrel stream, and at times nothing will push through; the issue was noticed prior to patient contact and there was no impact to any of the patient.A photo and a sample were received, several photos were provided, the bill of materials and production record were reviewed, and visual evaluation showed that several cannulas exhibited a forked or "double barreled" appearance.Based on review of the reported complaint, sample evaluation, photographic evidence, bill of materials, and production records, the reported issue was confirmed.The correction states, "this is a manufacturing process issue," and the pack build review found nothing that may have contributed to the reported defect, with the component received from the vendor and placed into the pack with no alteration by medline during the production process.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|