The reported issue was does not inflate, and the customer stated the bp cuff does not work.Medline received one compli-mates combination kit in used condition, tested the cuff according to the instructions for use, inflated the sample numerous times up to 300mmhg, observed no air leaks, confirmed the gauge was displaying readings as intended and the needle was moving freely, and conducted a pressure test using a bp simulator.The complaint was not confirmed, and the cause code was updated to reflect unconfirmed defect - not confirmed with received sample.Medline saw no issues during testing, and the sample worked as intended.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|