MED-EL Elektromedizinische Geraete GmbH SYNCHRONY; Cochlear Implant Back to Search Results
Model Number Mi1250 SYNCHRONY 2
Medical Device Problem Code Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
In-package measurement failed for electrode 3 in the operation room.
 
Event or Problem Description
In-package measurement failed for electrode 3 in the operation room.The device was not used and the user implanted with another device.
 
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Additional Manufacturer Narrative
Conclusion: the in-package measurements showed one channel affected with li impedance, thus a back-up implant was implanted.The reported issue was caused by a change of the telemetry measurement using maestro versions 10 or 11 and constitutes a false-fail measurement.Technically the implant works within specification and is fully functional.This is a final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRONY
Common Device Name
Cochlear Implant
Manufacturer (Section D)
MED-EL Elektromedizinische Geraete GmbH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
+43()577885626
MDR Report Key50771629
Report Number9710014-2026-00868
Device Sequence Number18747530
Product Code MCM
UDI-Device Identifier09008737395452
UDI-Public(01)09008737395452
Combination Product (Y/N)N
Initial Reporter CountryAR
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMi1250 SYNCHRONY 2
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/22/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2026
Supplement Date Received by Manufacturer09/02/2026
Initial Report FDA Received Date09/15/2026
Supplement Report FDA Received Date10/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/07/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-