| Model Number |
Mi1250 SYNCHRONY 2 |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
In-package measurement failed for electrode 3 in the operation room.
|
| |
|
Event or Problem Description
|
|
In-package measurement failed for electrode 3 in the operation room.The device was not used and the user implanted with another device.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the in-package measurements showed one channel affected with li impedance, thus a back-up implant was implanted.The reported issue was caused by a change of the telemetry measurement using maestro versions 10 or 11 and constitutes a false-fail measurement.Technically the implant works within specification and is fully functional.This is a final report.
|
| |
|