The account reported finding breaking hemostats.Medline received one hemostat as a sample for investigation; sample evaluation observed one jaw tip completely broke off from the rest of the component, and no other defects were noted with the rest of the hemostat.The reported problem was confirmed based on the sample evaluation, and the root cause is considered to be a vendor product issue relating to the component manufacturing process.This complaint has been provided to the component manufacturing facility for further evaluation, and the supplier quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|