The device was not returned for analysis.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information provided, the investigation determined that the reported issues and subsequent treatment appear to be related to operational context of the procedure.Factors that may contribute to difficulty removing the guide wire include, but are not limited to, inner diameter of the catheter, outer diameter of the guide wire, device support, buildup of procedural contaminants, challenging anatomy, and/or damage to the catheter or guide wire.In this case, it was reported that the guide wire was wedge under the stent struts which likely led to the reported stretched coils and tip separation.Based on the results of the complaint investigation, there is no indication a product quality issue with respect to manufacture, design or labeling of the device.
|