(b)(4).Date sent: 9/15/2026.D4 batch # 656d12.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.Photo analysis:
th is is an analysis of a photo submitted for evaluation.During the visual analysis, the following was observed:
the photo shows the shroud of device on the batch area (656d12 - 0074) and the part that support the anvil clamp doble pin was noted broken.Based on the photo the event described is confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the instrument was not returned our evaluation is limited.A manufacturing record evaluation was performed for the finished device batch number 656d12 , and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent:
could you please specify what was the component that was detached from the device?
was there any missing part of the anvil?
or, any missing part of the firing trigger?
or, any missing part from the handle?
if other, please specify.Is the current patient status known?
was there any patient consequence or change in the post-operative care of the patient as a result of the event? (extended hospital stay, readmission, re-operation, etc.).
|