A review of the complaint history for sn: (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn: (b)(6) was performed from the date of manufacture, 04-feb-2021 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the system had a door failure.A field service engineer found that the helmer pocket was not detecting when it was opened to dispense medication.The engineer shut down the device, replaced the irac board on the helmer unit, and then rebooted the device.After the replacement, the helmer operated normally and all functions were verified.The customer successfully recovered and inventoried the affected pockets, and no further issues were reported.The system functioned as intended after the field service engineer repaired the device.
|