Model Number 340-000-000
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Codes Respiratory Insufficiency (4462); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/14/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during rapid sequence induction for endotracheal intubation, the video laryngoscope exhibited a software malfunction resulting in a frozen/blocked screen.The device functioned normally during the pre-use check; however, the issue occurred at the time of intubation and was resolved after replacing the battery.A post-procedure test with the same battery reproduced the malfunction.The intubation was successfully completed, and no patient injury was reported.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
McGRATH
Common Device Name
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinnea road
memphis,tn
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinnea road
memphis,tn
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112--439
+1(800)646-4633
MDR Report Key50772486
Report Number3003120897-2026-01562
Device Sequence Number9425006
Product Code CCW
UDI-Device Identifier10884521824157
UDI-Public(01)10884521824157(17)270807(10)H25072104
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number340-000-000
Device Catalogue Number340-000-000
Device Lot NumberH25072104
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2026
Initial Report FDA Received Date09/15/2026
Date Device Manufactured08/07/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexFemale
-
-