| Model Number |
720157-01 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
Urinary Incontinence (4572)
|
| Date of Event |
07/20/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that after the six-week activation period for the artificial urinary sphincter (aus), the device did not activate and did not provide sufficient pressure for the cuff to coapt the urethra.As a result, the patient experienced urinary incontinence.Troubleshooting was performed, but it was unsuccessful.A cystoscopy was performed to visualize the cuff and verify that it was not providing sufficient pressure to close the urethra.A surgical procedure was performed, during which fluid loss was observed from a hole in the balloon; therefore, the balloon was removed and replaced.The original cuff and pump remained implanted.No further patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72404127.Serial number: n/a.Batch/lot number: 1100738086.Model/catalog description: control pump fgs iz.Unique identifier (udi) #: (b)(4).Model number/catalog number: 72400024.Serial number: n/a.Batch/lot number: 1100735463.Model/catalog description: balloon fgs 61-70 cm.Unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.
|
| |
|