The customer reported that the drape was opened according to standard practice, with some adherence of the plastic drape, and the adherence caused the previously opened portion to shift/move and change the location of the sterile field.Medline contacted the customer and requested additional information, reviewed the returned sample, sent the finished device alleged failure investigation form to the contract manufacturer, and performed analysis of complaint data; the sample was inspected and it was sticky and difficult to open, the contract manufacturer review found no abnormalities or defects in their production, and complaint data analysis determined that there is no trend.Based on the results of the investigation, medline was able to confirm the complaint issue.Medline was unable to determine the root cause of the issue with the returned sample, and concluded that no further investigation into the cause and complaint is warranted and that the complaint will be reopened, updated, and investigated as necessary if additional information becomes available.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|