Catalog Number NON25524
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/04/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the item is not sticking.
 
Additional Manufacturer Narrative
The customer called in stating the item is not sticking; the issue occurred during use and no injury occurred.The investigation noted that neither physical samples nor photos of the reported issue were provided by the customer; the vendor was notified of the issue, trend analysis was performed, and the vendor reviewed the production process, relevant records, and retained samples.The vendor noted that there were no indications of potential causes, and the coa was reviewed and determined that the product lot met all technical requirements including adhesive strength.Based on the information provided and the results of the investigation, medline was not able to confirm the reported issue or that the product failed to meet specifications; the cause was unconfirmed, sample required to determine cause, and no need for further action was indicated at this time.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
BANDAGE,ADHESVE,FABRIC,2"X4",ST,LF,50/BX
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
haroon mian
three lakes drive
northfield, IL 60093--275
MDR Report Key50772599
Report Number1417592-2026-40433
Device Sequence Number10614527
Product Code KGX
UDI-Device Identifier10080196325085
UDI-Public10080196325085
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberNON25524
Device Lot Number90924080001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2025
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-