A complaint was received related to weak adhesion from adhesive bandages; the customer stated that none of the bandages are sticking, and no injury was reported.Neither a physical sample nor photos of the reported issue were provided by the customer; the vendor¿s review of the production process, relevant records, and retained samples noted that there were no indications of potential causes, and retained samples were tested for peel strength and were noted to meet specifications.Based on the information provided and the results of the investigation, medline was not able to confirm the reported issue or that the product failed to meet specifications; the cause was documented as sample required to determine cause.If a sample, photo, and/or additional information becomes available, this complaint will be reopened to be further investigated.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|