| Model Number |
SIGADAPTSTND |
| Medical Device Problem Codes |
Entrapment of Device (1212); Retraction Problem (1536); Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
Tissue Breakdown (2681)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during a robotic rectal resection, a yellow error screen was displayed and the adapter was removed and waited until the error display disappeared from the screen, the adapter was then reconnected, but the error was not resolved.The tissue thickness during resection fluctuated between 1 and 2, and after the rectum was resected and the release button was pressed, the knife did not return, so the reload did not open.The monitor then displayed an error instructing removal of the cartridge.The adapter was removed, the error screen cleared, and the adapter was reconnected, but the issue was not resolved.The adapter and handle were then detached and reconnected again without response, and the green buttons were pressed and held for 10 seconds, but the knife still did not return.The articulation returned to straight when the button was pressed, but an attempt to return the knife using the manual adapter tool was unsuccessful because the tool was too hard and became bent.Ultimately, while the bowel remained clamped with the device, one 12 mm port was placed and another device was used to continue the approach, after which resection was completed without further issue.The proximal bowel was then resected through the small incision, and although the device could be removed, the cartridge remained clamped to the tissue.After removal from the body, an attempt was made to detach the device and reload, but they could not be detached; at that time, the blue lever remained in the lowered position and the red indicator remained displayed, although no red indicator was seen during the operation.Additional resection was performed, and additional suturing was completed with a stapler.
|
| |
|
Additional Manufacturer Narrative
|
|
D10: concomitant product sigphandle, sig power sigphandle handle, (serial no.(b)(6)), sigpshell, sig power sigpshell control shell, (lot no.#n5k1497y), egia60amt, egia60amt egia 60 artic med thick sulu, (lot no.#p5k1093), sigmret, sig power sigmret manual retract tool, (lot no.#c2016f077e).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|