| Model Number |
816300 |
| Medical Device Problem Code |
Inadequate User Interface (2958)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The field service representative (fsr) reported that during preventative maintenance (pm) of the device, the central control monitor (ccm) mouse on the touchscreen was off about half of an inch.Recalibration worked for a while, then it drifted again an inch off but was able to be recalibrated again and drifted again.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The fsr replaced the ccm and performed release testing.The unit operated to the manufacturer's specifications.
|
| |
|