• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; DBD-ARTHROSCOPY PACK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; DBD-ARTHROSCOPY PACK Back to Search Results
Catalog Number DYNJ64242B
Medical Device Problem Codes Moisture Damage (1405); Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "wet packs potentially used on patient.".
 
Additional Manufacturer Narrative
The account reported finding wet packs; more information stated the delivery person left some cases of this product outside while it was raining at the facility, the facility sent it back to medline, and the account received them again.No sample or photo was received for the investigation; finished good production records were reviewed and no non-conformances were noted, the finished good pack build was reviewed, trending was reviewed, and a stock check found no inventory left of this remaining lot at the branch.The reported problem could not be evaluated by sample or photo; based on the complaint details medline determined that the reported damage occurred after kit assembly during transit or storage of the pack.Given this the root cause is considered an operations service issue; per the facility the device was not used on a patient and therefore no patient involvement, and there was no serious injury identified, follow up care needed or medical intervention reported related to the incident.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
DBD-ARTHROSCOPY PACK
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
koushik pothamshetty
three lakes drive
northfield, IL 60093--275
MDR Report Key50773269
Report Number1417592-2026-40556
Device Sequence Number9471954
Product Code OJH
UDI-Device Identifier40193489956406
UDI-Public40193489956406
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNJ64242B
Device Lot Number22HMD098
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2023
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-