| Model Number |
BF-H190 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the bronchovideoscope had a separate issue.During the initial evaluation of the device it was observed, plastic distal end cover was burned.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's investigation.\
\
the device was returned to olympus for inspection and the reported event was confirmed.\
\
a definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: expected or random component failure without any design or manufacturing issue.\
\
should additional relevant information become available, a supplemental report will be submitted.\
\
olympus will continue to monitor field performance for this device.
|
| |
|