The customer reported that ¿2 kits are testing negative,¿ and the complaint details state ¿the positive control included with the kit tested negative a couple of times¿ and ¿most of the time, the positive control shows a very weak positive reaction (very light pink line).¿ at this time, ¿a sample(s) was not received for investigation,¿ and the investigation included ¿review of recent complaint trends/scar activity for the last three trend reports.¿ the complaint was not confirmed, and ¿without the sample or picture, it is difficult to truly identify the root cause of the issue.¿ no trends were identified for the specified material number and its corresponding defect type, and the division will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|