| Model Number |
TCDS0801-XTW |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026 a patient presented with grade 4 functional tricuspid regurgitation (tr) for a triclip procedure.During the procedure, triclip opened to about 15 degrees.Troubleshooting was performed and the clip was implanted.All steps were performed as per ifu.The final tr was grade <1.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.The investigation was unable to determine the cause for the reported unintended movement associated with clip open during efaa.There is no indication of a product issue with respect to manufacture, design, or labeling.
|
| |
|